• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, PEEK CORKSCREW FT; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, PEEK CORKSCREW FT; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SUTURE ANCHOR, PEEK CORKSCREW FT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Corrected information: d2a, d2b.Complaint allegation is confirmed.-visual inspection of the 4.5 mm peek corkscrew® ft suture anchor with one #2 fiberwire® and one # 2 tigerwire®, identified that the peek anchor was broken off at the tip and base, and the threads were damaged.-functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user error of the device due to misalignment insertion, and/or prying/leveraging of the device during insertion.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 2nd sep 2025, it was reported by an arthrex's employee via an email that for 1 unit of ar-1927psf-45, 4.5 mm peek corkscrew® ft suture anchor with one #2 fiberwire® and one # 2 tigerwire®, its anchor broke during insertion.This was detected during a rotator cuff repair surgery on (b)(6) 2025.The procedure was completed successfully by using another new ar-1927psf-45.There were no patient harm and no secondary procedure needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, PEEK CORKSCREW FT
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23095723
Report Number1220246-2025-03960
Device Sequence Number20987255
Product Code MBI
UDI-Device Identifier00888867023666
UDI-Public(01)00888867023666(11)240913(17)290831(10)15316978
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K061863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTURE ANCHOR, PEEK CORKSCREW FT
Device Catalogue NumberAR-1927PSF-45
Device Lot Number15316978
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2025
Supplement Date Received by Manufacturer12/03/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-