• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, BIOCOMPO- SITE PUSHLOCK; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, BIOCOMPO- SITE PUSHLOCK; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number SUTURE ANCHOR, BIOCOMPO- SITE PUSHLOCK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 2nd sep 2025, it was reported by an arthrex's employee via an email that for 1 unit of ar-1922bc, suture anchor, biocomposite pushlock® 4.5 x 24 mm, its anchor broke during insertion.This was detected during a shoulder instability surgery repair surgery on (b)(6) 2025.The procedure was completed successfully by using another new ar-1922bc.There were no patient harm and no secondary procedure needed.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.Visual inspection of the suture anchor, biocomposite pushlock® 4.5 x 24 mm, identified that the anchor was fractured at the midpoint of its length.The eyelet was not returned.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user error of the device due to misalignment insertion, and/or prying/leveraging of the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, BIOCOMPO- SITE PUSHLOCK
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23095756
Report Number1220246-2025-03961
Device Sequence Number21003946
Product Code MAI
UDI-Device Identifier00888867022874
UDI-Public(01)00888867022874(11)250212(17)290131(10)15388362
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K082810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTURE ANCHOR, BIOCOMPO- SITE PUSHLOCK
Device Catalogue NumberAR-1922BC
Device Lot Number15388362
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2025
Supplement Date Received by Manufacturer12/03/2025
Initial Report FDA Received Date09/19/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-