| Model Number |
SUTURE ANCHOR, PEEK CORKSCREW FT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 2nd sep 2025, it was reported by an arthrex's employee via an email that for 1 unit of ar-1927psf-45, 4.5 mm peek corkscrew® ft suture anchor with one #2 fiberwire® and one # 2 tigerwire®, the anchor broke during insertion.This was detected during a rotator cuff repair surgery on (b)(6) 2025.The procedure was successfully completed by using another new ar-1927psf-45.There were no patient harm and no secondary procedure needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Visual inspection of the 4.5 mm peek corkscrew® ft suture anchor with one #2 fiberwire® and one # 2 tigerwire®, identified that the peek anchor was broken off at the tip and the threads were damaged.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user error of the device due to misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|
Search Alerts/Recalls
|