Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The microsensor (id unknown) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history records (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.However, a probable root cause for the reported complaint could be an "unstable sensor performance".If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|
1 of 2 reports (same failure, different patients).Other mfg report number: 3013886523-2025-00255.A facility reported: "codman catheter did not produce readings on the cerelink monitor.In both situations, the patients had been under continuous monitoring for an extended period, and the issue did not appear to be related to transport or repositioning; the interruptions seem to occur spontaneously.This report is for the first patient/event: "the cerelink monitor stopped displaying both the waveform and the icp values.Troubleshooting included replacing the monitor and cables, but the issue persisted." "case required reassessment and intervention due to concerns about elevated icps, highlighting the critical impact of losing reliable monitoring.".
|