| Model Number |
GIF-H180J |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the gastrointestinal videoscope exhibited the light guide cover lens had foreign material.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation: the most probable cause of the reportable malfunction related to foreign material was not established, which indicates that the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.In addition, the following reportable malfunctions were identified during the device evaluation: the lens glue was chipped.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|