| Model Number |
9500-174 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an applicator deployment occurred.The sensor was inserted into the back of upper arm.Product was provided for investigation.An external visual inspection was performed and passed due to no damage.The internal visual inspection was performed and failed due detached needle.An applicator deployment was not found within the investigation window.The allegation was not confirmed.However, a broken or detached needle or cannula was found in connection to the reported event.The probable cause of the broken or detached needle or cannula could not be determined.No injury or medical intervention was reported.
|
| |
|