• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number DE2110X13B
Medical Device Problem Code Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging the active exhalation verification and motor calibration verification: reset motor calibration test steps had failed on a trilogy evo device.The device was not in use on a patient at the time of the occurrence.The trilogy evo device was returned to the third-party service center for evaluation.During the evaluation it was noted that the device's blower cable was not properly connected causing a proper connection to the system printed circuit assembly (pca) board and machine flow sensor had failed.The third-party service technician confirmed that it was the replaced machine flow sensor that failed for these test steps; not the one that had contamination on it.In conclusion, the device's blower cable was re-connected to the system pca board, and the machine flow sensor was replaced to address the issues.The root cause is confirmed at this time.Additionally, during the service of the device, contamination found on the blower assembly, machine flow sensor, and in-line filter proximal which were unrelated to the reportable malfunction/issue.The following parts were replaced as part of the four-year preventative maintenance on this device: muffler assembly, air foam inlet filter, and particulate filter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23096833
Report Number2518422-2025-110388
Device Sequence Number9226304
Product Code CBK
UDI-Device Identifier00606959054110
UDI-Public(01)00606959054110
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDE2110X13B
Device Catalogue NumberDE2110X13B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2025
Initial Report FDA Received Date09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-