There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Insert received has a different manufacture date and different tip lot number than the insert in the case.Return qty (b)(4), 30k thinsert fg-73, 00131734, bul per manufacture date.Test method / gp005-wi02.Inductance: gauge id# 5349 due: 12/31/2026, result 3.14.Meets spec, does not meet spec, n/a.Tip condition: - meets spec, does not meet spec, n/a.Stack condition, meets spec, does not meet spec, n/a.Tested on: g130 gage id# 9626-2, due date: 03/31/26, set pressure: 35 psi.Water flow: output rate: 35 psi - meets spec, does not meet spec, n/a.Spray: - meets spec, does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring, seam leak, n/a.Vibration: - meets spec, does not meet spec, n/a.Grip condition: meet, does not meet spec, n/a.Other visual observation: insert is good, no fault found.Insert was tested on a digital thermometer i.D.#6116a.Due date:09/30/2026.The temperature was 79.2 °f, no fault found with temperature.The specifications state the temperature is not to exceed 118.4°f.(per (b)(4)) alleged event: confirmed, not confirmed.
|