| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
09/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.A decrease in hearing was observed over the past month and it was reported that the sound spontaneously increases and decreases from time to time.Re-implantation is being considered.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered, but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, an impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|