This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 10, 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual inspection was performed.The implant had signs of use with debris on its external threads.During a functional test the mount disengaged from the implant as intended.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did submit one image of the reported event.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was possibly user caused.However, during functional testing the product functioned as intended, no problem found.Therefore, based on the available information, device malfunction has not occurred.During a functional test the mount disengaged from the implant as intended.The reported event was unconfirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|