| Catalog Number |
82250240S |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported; "miss target occurred during drilling for inserting the screw to the distal oval hole of gamma4 nail.The screw was once removed, freehand drilling was performed, and screw was reinserted.The surgery was completed successfully after that while there was about 10 minutes delay.The surgeon is aware of the fact that there is no problem with the devices because a functional check of the nail and the targeting device was performed prior to insertion of the nail into the patient's body.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the affected device was not returned and no evidence was provided for evaluation.A device inspection was not possible since the affected device was not returned.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.Indications of material, manufacturing or design related problems were unable to be identified as the lot number was not communicated.A review of the labelling did not indicate any abnormalities.More information, as well as the affected device, must be available in order to determine the exact root cause of the issue.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported; "miss target occurred during drilling for inserting the screw to the distal oval hole of gamma4 nail.The screw was once removed, freehand drilling was performed, and screw was reinserted.The surgery was completed successfully after that while there was about 10 minutes delay.The surgeon is aware of the fact that there is no problem with the devices because a functional check of the nail and the targeting device was performed prior to insertion of the nail into the patient's body.".
|
| |
|