| Model Number |
CD2411-36C |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
It was reported that atrial oversensing was noted on the implantable cardioverter defibrillator (icd) and right atrial (ra) lead.No changes or intervention was reported.The patient condition was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Section g2 - "healthcare professional" should not be selected.
|
| |
|
Event or Problem Description
|
|
New information received notes that the oversensing resulted in pacing inhibition.Programming changes were made to resolve the issue.There were no patient consequences noted.
|
| |
|