The product investigation was completed.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Visual inspection and screening test of the returned device were performed following j&j medtech procedures.Visual analysis revealed a hole and reddish material inside the pebax.The force feature was tested and no errors were observed.The force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device number lot 31559779l and no internal actions related to the complaint were found during the review.The reddish material inside the pebax could be related to the force issue reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this could not be conclusively determined.The instructions for use (ifu) of the carto 3 system contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter; the instruction for use of the device contain the following recommendations: zero the contact force reading following insertion of the catheter into the patient.The tip electrode and all four ring electrodes on the catheter tip must be outside of the sheath so that the force sensor is inside the body.Variations in the force reading at the same rate as the cardiac or respiration cycle may indicate contact with cardiac structures.When these markers indicate the catheter tip is not in contact, the reading can be zeroed.Do not scrub or twist the tip electrode because damage to the tip electrode bond may occur and loosen the tip electrode or damage the contact force sensor and affect measurement accuracy.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|