| Catalog Number |
0380544100 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the shaver broke off.They were able to remove all the parts from the joint.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the tip of the shaver broke off.They were able to remove all the parts from the joint.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an excessive force applied to the tip that will contribute to the failure.Inadequate welding process.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the shaver broke off.They were able to remove all the parts from the joint.
|
| |
|