MEDOS INTERNATIONAL SÃ RL TRUESPAN 0 DEGREE PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
228150 |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a meniscal repair surgical procedure it was observed that the truespan 0 degree peek device red trigger was broken during insertion of the implant.This occurred two times with different guns.The procedure was completed successfully.It was unknown if there was a delay in the procedure.There were fragments generated.The fragments were removed.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.This is report 1 of 2 of (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device, and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|