| Catalog Number |
0165L20 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(6) hospital.The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that spontaneous removal of foley catheter was discovered on the second day after placement.Observation of the cuff revealed damage.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(6) hospital.The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: d.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that spontaneous removal of foley catheter was discovered on the second day after placement.Observation of the cuff revealed damage.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ cause unknown.The reported failure was able to be reproduced.Visual inspection observed irregular burst without missing pieces.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.However, the exact cause on how and when problem occurred could not be determined.A potential root cause could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ thin sac).The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Initial reporter facility name: (b)(6) hospital.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that spontaneous removal of foley catheter was discovered on the second day after placement.Observation of the cuff revealed damage.
|
| |
|