| Catalog Number |
VSTL45 |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H6.Component code: g07002 - device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did this issue contribute to any patient adverse event? this report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported to the sales rep that on 9/19/2025 the customer received four boxes, each with the correct product code label.However, the items inside the boxes were incorrect.The customer ordered product code for the laparoscopic dual flexible applicator, but the actual items received were laparoscopic dual rigid applicator.The boxes themselves are correctly labeled, but the contents do not match the ordered product, resulting in the correct boxes containing the wrong items.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Corrected information: d 9.Device available for evaluation? additional information: d 9.Date device returned to manufacturer, d 9.Is device returned to manufacturer?, h 6.Component code, h 6.Type of investigation, h 6.Investigation findings, h 6.Investigation conclusions.H6 component code: g07002 - pending evaluation of returned device.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Corrected information: d 4.Primary udi number, h 3.Device evaluated by manufacturer? additional information: d 4.Expiration date, h 4.Device manufacture date, h 6.Type of investigation, h 6.Investigation findings, h 6.Investigation conclusions, h 6.Component code.H3 investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the secondary box of the vstl45 product code with lot: 5881559 was returned.Upon evaluation of the packaging inside of the secondary box, the primary packaging of product code vstl35 with the re-label of lot: 5881559 was found inside.No conclusion could be reached as to what may have caused the reported incident.The reported complaint was confirmed.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A review of the device batch record was performed for raw materials, in-process and finished goods for lot number 5881559.Capa-03781 was opened and is currently in voe phase.The devices met all raw materials, in-process and finished goods specifications upon release of the product.This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|