Catalog Number CS-24301-E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "22 aug 2025, the doctor found a significant amount of gas can still be aspirated 2 cm below the skin surface, suggesting the needle is unable to maintain negative pressure." the user changed to a new set to resolve the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Event or Problem Description
It was reported that "(b)(6) 2025, the doctor found a significant amount of gas can still be aspirated 2 cm below the skin surface, suggesting the needle is unable to maintain negative pressure." the user changed to a new set to resovle the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 16 GA X 8IN (20 CM)
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23132813
Report Number3006425876-2025-00879
Device Sequence Number18354709
Product Code FOZ
UDI-Device Identifier00801902102225
UDI-Public(01)00801902102225(11)240814(17)260702(10)71F24H0287(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/02/2026
Device Catalogue NumberCS-24301-E
Device Lot Number71F24H0287
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2025
Supplement Date Received by Manufacturer10/15/2025
Initial Report FDA Received Date09/24/2025
Supplement Report FDA Received Date10/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED