| Model Number |
9-ATV08F45/80 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
09/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 9mm amplatzer septal occluder was chosen for implant using an 8f amplatzer trevisio intravascular delivery system.During the procedure, the cobra device deformation occurred while expanding the device within the patient's anatomy.The device was removed from the patient's anatomy prior to being released from the delivery cable.A new 9mm amplatzer septal occluder was chosen using the same delivery system, a cobra device deformation occurred again.The deformed device were retrieved outside the patient prior to release from the delivery cable.It was observed that a thrombus had developed on the 8f delivery system.The thrombus did not have unusual or fiber like appearance.It was noted that the sheath was in the patient for a significant period of time due to the deformations.The activated clotting time (act) level was noted as 250 during the procedure, however it was noted that it was difficult to prolong and additional heparin was administered.The patient did not have any hematological disorders or systemic illness to contribute to the hypercoagulability.The procedure was completed using an 8mm amplatzer septal occluder and new deliver system.It was reported that there was no clinically significant delay throughout the procedure and the patient remained hemodynamically stable throughout the procedure.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of thrombus formation observed on the delivery sheath was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on the available information, the cause of the reported thrombus formation could not be conclusively determined.It is possible that the unplanned prolonged procedure contributed to the reported incident; however, this cannot be confirmed.The reported serious injury/illness/impairment was a result of case-specific circumstances as no intervention was reported.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Codes removed: health effect- impact code: 4614 serious injury/ illness/ impairment.Correction: d4 - lot #, d4 - expiration date, h4 - device mfg date.
|
| |
|
Search Alerts/Recalls
|