| Model Number |
CR2 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer has been sent a replacement device.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not power on when the device lid was activated.In this state defibrillation therapy may be delayed if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that the cause of the reported issue was due to the lid switch, designator sw5.Stryker archived the device.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not power on when the device lid was activated.In this state defibrillation therapy may be delayed if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use reported with the event.
|
| |
|