| Catalog Number |
INTP14 |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported by customer that the first pulled the straight catheter tray paper back the catheter was exposed and not within the sterile kit.The nurse said it¿s happened with 2 to 3 kits lately and stated they have had the same issue at least 7 times within the last 2 weeks.One nurse said the catheter tip fell out as soon as the package was pulled from the plastic sleeve.Per customer follow up information received via email on 23sept2025, it did occur again on the 17th in the ed with a straight catheter kit.They discarded the kit and provided the material ref intp14 and lot # ngku0148.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings, and no sample was available for evaluation.A labeling review was not performed because labeling could not have prevented the reported failure.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: d, e.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported by customer that the first pulled the straight catheter tray paper back the catheter was exposed and not within the sterile kit.The nurse said it¿s happened with 2 to 3 kits lately and stated they have had the same issue at least 7 times within the last 2 weeks.One nurse said the catheter tip fell out as soon as the package was pulled from the plastic sleeve.Per customer follow up information received via email on 23sept2025, it did occur again on the 17th in the ed with a straight catheter kit.They discarded the kit and provided the material ref intp14 and lot # ngku0148.
|
| |
|
Search Alerts/Recalls
|