• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO JUNCOS RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO JUNCOS RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 37714
Medical Device Problem Codes Break (1069); Connection Problem (2900)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the manufacturer representative (rep) reports that the doctor contacted him on friday to report that they did an ins replacement on (b)(6) 2023.Dr reports that they had trouble removing the lead from the old ins and then they had trouble inserting the lead into the new ins.Dr thinks the lead contact was "crushed".Dr reports that they were unable to insert the lead all the way into the new ins and the last contact is outside the ins header.Rep said the dr is not planning on replacing the lead.Rep did not have the patient (pt) information at time of call to technical services (ts).
 
Additional Manufacturer Narrative
Section d references the main component of the system.Other medical products in use during the event include: brand name lead; product id neu_unknown_lead (serial: unknown); product type: 0200-lead; medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Continuation of d10: product id 39565-65 serial# (b)(6) implanted: (b)(6) 2013 product type lead product id 977006 serial# (b)(6) implanted: (b)(6) 2025 product type implantable neurostimulator medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received from the manufacturer representative (rep).It was reported that the cause of the crushed electrodes was unknown.Physician attempted to restore, crushed contact to allow passage into ins.Physician was unable to restore, damaged contact.They were remains an exposed contact outside of the battery.Patient was made aware of the situation.No further intervention is planned by the physician at this time.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The rep reported the previous ins was discarded by the facility per their policy, and no return will be possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE SENSOR
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO JUNCOS
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO JUNCOS
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key23144728
Report Number2182207-2025-02560
Device Sequence Number9284624
Product Code LGW
UDI-Device Identifier00613994610430
UDI-Public(01)00613994610430(17)131228
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2013
Device Model Number37714
Device Catalogue Number37714
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2025
Supplement Date Received by Manufacturer11/14/2025
12/02/2025
Initial Report FDA Received Date09/25/2025
Supplement Report FDA Received Date11/17/2025
12/03/2025
Date Device Manufactured02/05/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11....".; SEE H11...
Patient Age78 YR
Patient SexMale
-
-