The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The investigation did not identify any conditions or factors within the product that could have contributed to the reported event.Visual: received one (1) used all silicone foley catheter without original packaging.Verified material number 119114 and manufacturing lot number ngjs4653.Visual inspection noted no obvious observations.Using the returned syringe, the catheter balloon was inflated with 10ml methylene blue solution (3 drops 1 percent aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.Asymmetrical balloon observed at 80:20.The balloon passively deflated in under two (2) minutes with no cuffing or leaks noted, returning 10ml of solution.This is within specification per inspection procedure which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed, a risk and labeling review is not required.A review of the device history record was not necessary due to the reported event being unconfirmed.Based on the investigation results no additional action is required at this time.Corrections: d, e, f, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during use, foley catheter leaked at y down straight.Patient was exposed to urine.No medical intervention was reported.Per notification from ucc on 25dec2025, it was reported that asymmetrical balloon was found during sample evaluation on 05nov2025.
|