MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER Back to Search Results
Catalog Number 119114
Medical Device Problem Codes Material Puncture/Hole (1504); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Exposure to Body Fluids (1745)
Date of Event 09/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use, foley catheter leaked at y down straight.Patient was exposed to urine.No medical intervention was reported.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The investigation did not identify any conditions or factors within the product that could have contributed to the reported event.Visual: received one (1) used all silicone foley catheter without original packaging.Verified material number 119114 and manufacturing lot number ngjs4653.Visual inspection noted no obvious observations.Using the returned syringe, the catheter balloon was inflated with 10ml methylene blue solution (3 drops 1 percent aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.Asymmetrical balloon observed at 80:20.The balloon passively deflated in under two (2) minutes with no cuffing or leaks noted, returning 10ml of solution.This is within specification per inspection procedure which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed, a risk and labeling review is not required.A review of the device history record was not necessary due to the reported event being unconfirmed.Based on the investigation results no additional action is required at this time.Corrections: d, e, f, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during use, foley catheter leaked at y down straight.Patient was exposed to urine.No medical intervention was reported.Per notification from ucc on 25dec2025, it was reported that asymmetrical balloon was found during sample evaluation on 05nov2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23145090
Report Number1018233-2025-08585
Device Sequence Number12972356
Product Code EZL
UDI-Device Identifier00801741046025
UDI-Public(01)00801741046025(11)240726(17)290430(10)NGJS4653
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number119114
Device Lot NumberNGJS4653
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2025
Supplement Date Received by Manufacturer04/07/2026
Initial Report FDA Received Date09/25/2025
Supplement Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-