| Catalog Number |
SRX-800 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
08/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist abbvie in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade iv" this record is for the right side.The device was explanted and replaced.Device was returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases, wear abrasion and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|