| Catalog Number |
AE05ML |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "the ae05ml product was used during a lung lobe resection procedure.Due to the fact that only one clip came out while the rest were stuck and couldn't be removed, this incident did not cause any injuries to personnel".No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) the customer returned one unit of ae05ml auto endo5 ml for investigation.The returned sample was visually examined with and without magnification.Visual examination of the returned sample revealed a broken safety pin in the safety pin hole.The hole where the safety pin is placed was deformed on the applier's handle frame.There was no biological material observed on the device.The reported complaint of "broken safety pin" was confirmed based upon the sample received.The device was returned with a broken safety pin and damaged the safety pin hole.Proper functional testing could not be completed because the applier was unable to be engaged due to the broken safety pin.R & d concluded the probable root cause for the safety pin breakage is a fracture defect with the supplied safety pins.A device history record review was performed, and no relevant findings were identified.Actions to address the safety pin break were established as part of a non-conformance, which was closed on 29-jul-2025.The returned sample was manufactured prior to these actions on 15-mar-2025.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "the ae05ml product was used during a lung lobe resection procedure.Due to the fact that only one clip came out while the rest were stuck and couldn't be removed, this incident did not cause any injuries to personnel".No patient harm or injury.The patient status is reported as "fine".
|
| |
|