• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALTER LABS DE MEXICO DE S.A. DE C.V. PARKER; PARKER FLEX-TIP TRACHEAL TUBE WITH STYLET, HIGH VOLUME LOW PRESSURE, CUFFED (PFS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALTER LABS DE MEXICO DE S.A. DE C.V. PARKER; PARKER FLEX-TIP TRACHEAL TUBE WITH STYLET, HIGH VOLUME LOW PRESSURE, CUFFED (PFS Back to Search Results
Model Number I-PFST-70
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has not been returned for evaluation.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided.All information reasonably known as of 25 sep 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported, the pilot balloon deflated immediately after intubation, (there was no problem with the cuff test prior to intubation.) the patient was reintubated, there was no reported injury.
 
Event or Problem Description
It was reported, the pilot balloon deflated immediately after intubation, (there was no problem with the cuff test prior to intubation.) the patient was reintubated, there was no reported injury.
 
Additional Manufacturer Narrative
Correction: d9 (sample return).Additional information-investigation completion: h2; h6 functional testing was performed on the returned product, by the distributor; analysis of the device revealed that the pilot balloon remained inflated; therefore, the reported issue was not confirmed.A review of the device history record is not possible as no lot number was provided.A risk assessment was performed, and the ultimate risk was determined to be low.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided.All information reasonably known as of 04 mar 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a sun med holdings llc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARKER
Common Device Name
PARKER FLEX-TIP TRACHEAL TUBE WITH STYLET, HIGH VOLUME LOW PRESSURE, CUFFED (PFS
Manufacturer (Section D)
SALTER LABS DE MEXICO DE S.A. DE C.V.
blvd. independencia #2167
parque ind. las americas
horizonte sur. cd. juarez, mx 32596
MX  32596
Manufacturer (Section G)
SALTER LABS DE MEXICO DE S.A. DE C.V.
blvd. independencia #2167
parque ind. las americas
horizonte sur. cd. juarez, mx 32596
MX   32596
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 
MDR Report Key23151445
Report Number3004748541-2025-00067
Device Sequence Number10102629
Product Code BTR
UDI-Device Identifier00607411958014
UDI-Public(01)00607411958014
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K984528
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberI-PFST-70
Device Catalogue NumberI-PFST-70
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2025
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date09/25/2025
Supplement Report FDA Received Date03/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-