| Model Number |
7735 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
09/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead was part of a system revision due to infection and vegetation on lead.There was no additional adverse patient effects were reported.This ra lead was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against this device was not confirmed.This device was analyzed, passed all the baseline tests and exhibited normal device functions.This supplemental is being filed to capture d9: returned to manufacturer date and investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead was part of a system revision due to infection and vegetation on lead.There was no additional adverse patient effects were reported.This ra lead was explanted.
|
| |
|