| Catalog Number |
8065772265 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported that ophthalmic knives from four different lot numbers were found to be blunt during a cataract surgery.The procedure was successfully completed on the same day by using a replacement knife, and there was no impact to the patient.No further information is expected, as the customer was unwilling to provide additional details.This report pertains to the third case, involving one unknown lot numbers.A total of 15 blunt knives were reported across three cases.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections d.4.D.9.H.3.And h.4.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections h.3, h.6 and h.11.One opened knife was received for the report of blunt knives.Sample was visually inspected and found to be nonconforming with a damaged cutting edge.Penetration testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The returned sample is visually non-conforming with a damaged cutting edge.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint sample could have contributed to the customers reported issue of blunt knives.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All knives are 100% visually inspected by trained operators using a minimum of 10x magnification during manufacturing.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|