• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0636
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that once the capsule was placed, it was checked with an endoscope.The capsule had moved, coming loose from its anchoring location, and was found in the patient's esophagus.Another capsule was used during the same procedure, and it attached successfully.There was no patient or user harm, and no intervention was required.There was nothing unusual about the patient or the procedure.
 
Event or Problem Description
It was reported that once the capsule was placed, it was checked with an endoscope.The capsule had moved, coming loose from its anchoring location, and was found in the patient's esophagus.Another capsule was used during the same procedure, and it attached successfully.There was no patient or user harm, and no intervention was required.There was nothing unusual about the patient or the procedure.No lubricant was used to facilitate the placement of the capsule.
 
Additional Manufacturer Narrative
Additional information: b5, d4(udi), g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Common Device Name
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23154338
Report Number9710107-2025-00432
Device Sequence Number12056832
Product Code FFT
UDI-Device Identifier07290101369714
UDI-Public(01)07290101369714(17)261216(10)65922F(20)01
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K102543
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFGS-0636
Device Catalogue NumberFGS-0636
Device Lot Number65922F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2025
Supplement Date Received by Manufacturer10/28/2025
Initial Report FDA Received Date09/26/2025
Supplement Report FDA Received Date11/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-