Catalog Number 823162
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/29/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported a hakim valve (id 823162) was implanted due to congenital hydrocephalus via ventriculoperitoneal (vp) shunt on unknown date with unknown setting.The valve was removed and replaced on (b)(6) 2025, due to suspicion of obstruction.According to the information provided, it is unknown if the patient experienced any signs and symptoms due to product problem.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Updated fields: g3, g6, h11.Root cause analysis - no root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the reported issue could be due to biologicals debris, as mentioned in the ifu, accumulation of biological matter (i.E.Blood, protein accumulations, tissue fragments, etc.) in the programming mechanism can cause inability of the device to be reprogrammed.Clogging of the programmable valve with biological matter can cause the valve to become unresponsive to attempts to change the pressure setting.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The hakim valve (id 823162) was returned for evaluation.Device history record (dhr) - the product code 82-3162 with lot 7356566 conformed to the specifications when released to stock.Failure analysis - the position of the cam when the valve was received was at setting 110 mmh2o.The valve was visually inspected, no defect was noted.The valve passed the test for programming, occlusion, leak, reflux, siphon guard and pressure.Root cause analysis - no root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the reported issue could be due to biologicals debris, as mentioned in the ifu, accumulation of biological matter within the valve can cause difficulties adjusting the valve setting with the tool kit, and impair the anti-reflux function.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA 02048
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA 02048
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key23154787
Report Number3013886523-2025-00258
Device Sequence Number14775086
Product Code JXG
UDI-Device Identifier10381780519232
UDI-Public(01)10381780519232(17)290228(10)7356566
Combination Product (Y/N)N
PMA/510(K) Number
K221840
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number823162
Device Lot Number7356566
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2025
Supplement Date Received by Manufacturer11/07/2025
12/11/2025
Initial Report FDA Received Date09/26/2025
Supplement Report FDA Received Date11/26/2025
01/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
Patient SexFemale
-
-