|
It was reported to boston scientific corporation that a truetome 44 was attempted to be used in the bile duct during a procedure performed on (b)(6) 2025 during the procedure, the cutting wire of the truetome 44 broke inside the patient's bile duct.It was reported that a portion the cut wire broke at approximately 5 cm, and they were able to remove the device without it being free in the cavity.No further information has been obtained despite good faith efforts.There were no reported patient complications as a result of this event.
|
|
It was reported to boston scientific corporation that a truetome 44 was attempted to be used in the bile duct during a procedure performed on (b)(6) 2025 during the procedure, the cutting wire of the truetome 44 broke inside the patient's bile duct.It was reported that a portion the cut wire broke at approximately 5 cm, and they were able to remove the device without it being free in the cavity.No further information has been obtained despite good faith efforts.There were no reported patient complications as a result of this event.
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: (investigation result).The returned truetome 44 was analyzed, and a visual evaluation noted that the cutting wire was blackened, broken, and detached.The working length was also torn from the distal pierce hole.Additionally, the wire anchor was dislodged from the working length and the tip was damaged.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken and detached.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The break in the cutting wire could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity, causing premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can detach the cutting wire as observed in the product analysis.The working length torn at the pierce hole, could have been caused by submitting the cutting wire to tension during the handle actuation, in addition with the possibility of the device being energized during the handle actuation.Also bowing the device without being completely out of the scope can lead to tear it and displacing or dislodge the cutting wire anchor from its position.Additionally, the damaged tip could be caused by operational factors, such as manipulating the device through difficult anatomy, the used technique for the device manipulation, the interaction between the truetome and the scope (hitting against a hard surface) or the insertion and removal of the guidewire or by the mandrel removal.Based on all gathered information, the most probable root cause is adverse event related to procedure.
|