• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850AEA
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).We have requested the return of the subject mr850 respiratory humidifier speaker for investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A distributor in china reported on behalf of a healthcare facility that the audible alarm of the mr850 respiratory humidifier was not functioning.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Additional Manufacturer Narrative
(b)(6).Method: the subject mr850 respiratory humidifier was returned to fisher & paykel (f&p) healthcare regional office in china where it was inspected by a trained f&p healthcare service technician.Our investigation is based on the information provided by the f&p healthcare service technician, information provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: performance testing of the mr850 respiratory humidifier confirmed that no sound was generated from the speaker, confirming the reported issue.Conclusion: without the return of the subject device, or the speaker, we are unable to determine the root cause of the reported issue.The mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
A distributor in china reported on behalf of a healthcare facility that the audible alarm of the mr850 respiratory humidifier was not functioning.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23164009
Report Number9611451-2025-00959
Device Sequence Number8882486
Product Code BTT
UDI-Device Identifier09420012407177
UDI-Public(01)09420012407177(10)2102128940(11)220408
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K073706
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR850AEA
Device Catalogue NumberMR850AEA
Device Lot Number2102128940
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2025
Supplement Date Received by Manufacturer10/23/2025
Initial Report FDA Received Date09/28/2025
Supplement Report FDA Received Date11/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-