C.R. BARD, INC. (BASD) -3006260740 POWERPORT M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
1808550 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Difficult to Flush (1251); Fracture (1260); Material Separation (1562); Stretched (1601); Material Protrusion/Extrusion (2979); Migration (4003)
|
| Health Effect - Clinical Codes |
Extravasation (1842); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022.A patient underwent a port placement procedure using powerport m.R.I.Implantable port.Sometimes later post port placement, it was reported that the catheter was allegedly experienced issues with flushing.It was further reported that the catheter was found to be fractured.Reportedly, the catheter was migrated into patient's heart.The current status of the patient was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection / evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device / patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: although the physical device was not returned for evaluation, two images were provided for review.One image shows device placed via a right jugular access.Contrast is being infused, extravasation is noted at the arch of the catheter in the neck.Second image shows the device introduced via a right jugular access is unremarkable.The investigation is confirmed for the reported catheter fracture, difficult to flush and fluid leak issues.However, it is inconclusive for the reported stretched and material protrusion or extrusion issues as exact circumstances of the reported event cannot be verified from images.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, d6a, d6b, g3, h6 (device, component, method, result) section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022.A patient underwent a port placement procedure using powerport m.R.I.Implantable port.Sometimes later post port placement, on (b)(6) 2025, it was reported that post port placement, the catheter was allegedly experienced issues with flushing.It was further reported that the catheter was allegedly found fractured.Reportedly, patient allegedly experienced that the catheter migrated to the heart.Reportedly, patient allegedly experienced pain when catheter was flushed and a grape sized bump along port catheter, contrast and saline extravasation occurred along the catheter, the current status of the patient was unknown.
|
| |
|