| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after sustaining a head trauma.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered, but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after sustaining a head trauma.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after sustaining a head trauma.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation was carried out but the device has not been received yet for investigational purposes.
|
| |
|