• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 50MM D; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 50MM D; PROSTHESIS, HIP Back to Search Results
Catalog Number 110010263
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 09/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: cat# 30123204 lot# 66963690 32mm i.D.Size d high wall liner.Cat# 00625006525 lot# j7633718 bone screw self-tapping 6.5 mm dia.25 mm length.Cat# 00625006515 lot# 64596056 bone screw self-tapping 6.5 mm dia.15 mm length.Cat# 00625006520 lot# j7643022 bone screw self-tapping 6.5 mm dia.20 mm length.Cat# 00625006530 lot# j7900452 bone screw self-tapping 6.5 mm dia.20 mm length.Cat# 00625006515 lot# j7458207 bone screw self-tapping 6.5 mm dia.15 mm length.Cat# 00625006520 lot# j7617986 bone screw self-tapping 6.5 mm dia.20 mm length.Unk head.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial hip procedure.Subsequently, the patient was revised approximately two months later due to dislocation and cup malposition.The cup was retroverted.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; g3; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no deviations or anomalies during manufacturing.Total hip arthroplasty.Acetabular cup abnormal steep lateral angle of inclination is a risk factor for hip dislocation.Cannot exclude loosening of the acetabular cup and osteolysis of the acetabular roof, as reported above.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 50MM D
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23166837
Report Number0001825034-2025-03033
Device Sequence Number13164934
Product Code LPH
UDI-Device Identifier00887868356018
UDI-Public(01)00887868356018(17)330503(10)7499300
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number110010263
Device Lot Number7499300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2025
Supplement Date Received by Manufacturer12/23/2025
Initial Report FDA Received Date09/29/2025
Supplement Report FDA Received Date12/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexUnknown
Patient Weight90 KG
-
-