| Model Number |
NVTR-27 |
| Medical Device Problem Codes |
Off-Label Use (1494); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
09/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 27mm navitor valve was chosen for implantation utilizing a large flexnav delivery system.Aortic angle was 63 degrees.Sufficient calcium present for anchoring.Patient presented with bicuspid aortic valve, significant leaflet calcification, and a calcium block between the left and right sinuses.Pre-dilatation was performed with a 22mm balloon.Navitor valve was advanced.At 80% deployment, the valve was under expanded at a depth of 5mm.Under expansion thought to be due to excess calcium and bicuspid anatomy.It was decided to release in a lower position with a planned post dilatation.After deployment at 3mm the valve migrated towards aorta.All tabs had released and guidewire was at midventricular position.Post dilatation was performed but following this, developed aortic insufficiency and cardiac arrest.Chest compressions were performed.A second 27mm navitor valve was then implanted valve in valve with a new large flexnav, but the patient remained in cardiac arrest and chest compressions continued.It was then decided to place the patient on cardiopulmonary bypass (cbp).After this, it was discovered that the left coronary artery was partially obstructed by the first valve.A stent was placed and the patient began showing improvement.The procedure concluded and patient was transferred to icu with ecmo support.Left ventricular function normalized, the valve was functioning normally with a gradient of 13mmhg.Physician thought the migration was due to the bicuspid anatomy.
|
| |
|
Event or Problem Description
|
|
Subsequent to the previously filed report, additional information was received the patient was reported to now be off ecmo and progressing well.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of cardiac arrest, and off-label use was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the available information, the cause of the reported cardiac arrest could not be conclusively determined.The reported off-label use was related to the use of the navitor on a bicuspid aortic valve.The reported surgical and unexpected medical intervention were results of case-specific circumstances as resuscitation was performed due to cardiac arrest, cardiopulmonary bypass was initiated, and the patient was eventually transferred to the icu under ecmo.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Codes removed: medical device problem code: 2993 adverse event without identified device or use problem.Codes added: medical device problem code: 1494 off - label use.
|
| |
|
Search Alerts/Recalls
|