| Model Number |
9-ASD-030 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 30mm amplatzer septal occluder was chosen for implant using a 12f non-abbott device.During the procedure, the physician noticed a cobra formation on the left disc and the issue persisted even after multiple attempts.The case was postponed to a later date.It is noted that the deformed device was never released from the delivery cable while inside the patient.There was no interaction with the cardiac structures during deployment and no angulation or kink noticed in delivery system.It was observed that there was no clinically significant delay and the patient remained hemodynamically stable throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|