| Catalog Number |
72201993 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a shoulder joint labrum repair surgery, it was noticed that the osteoraptor anchor broke while being implanted.All the pieces were removed with grasping forceps and suction.The procedure was resumed, after a non-significant delay using an equivalent s+n back-up.The back up was used in the originally drilled bone.There was no void left in the patient and the instrument to prepare the insertion site was a 2.3mm anchor matching instrument.The insertion site was cleared of all debris prior to insertion no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder joint labrum repair surgery, it was noticed that the osteoraptor anchor broke while being implanted.All the pieces were removed with grasping forceps and suction.The procedure was resumed, after a non-significant delay using an equivalent s+n back-up.The back up was used in the originally drilled bone.There was no void left in the patient and the instrument to prepare the insertion site was a 2.3mm anchor matching instrument.The insertion site was cleared of all debris prior to insertion no further complications were reported.Patient status is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H11 h3, h6: part of the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The anchor was not returned.One blue/white suture was returned.The insertion device has no visible defect.A material assessment could not be performed due to the part not being returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include an inadvertent impact event that may have occurred during device transportation, rough shipping and handling, or at the customer site, as well as the application of an unintended, inappropriate, or excessive force to the device.Based on this investigation, the need for corrective action is not indicated.
|
| |
|