| Catalog Number |
628516S |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Erythema (1840); Skin Inflammation/ Irritation (4545); Swelling/ Edema (4577)
|
| Date of Event |
09/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "stimulation of soft tissue due to locking screw backout.The locking screw was removed." 09/06/2025 - additional information received: "the screw at the most distal end of the hook plate was also loose, but that was completely locked when it was re-tightened, so it was re-tightened.The backed-out screw caused irritation, and the skin on the clavicle surface was red and swollen.".
|
| |
|
Event or Problem Description
|
|
As reported: "stimulation of soft tissue due to locking screw backout.The locking screw was removed." 09/06/2025 - additional information received: "the screw at the most distal end of the hook plate was also loose, but that was completely locked when it was re-tightened, so it was re-tightened.The backed-out screw caused irritation, and the skin on the clavicle surface was red and swollen.".
|
| |
|
Additional Manufacturer Narrative
|
|
A device inspection was not possible since the affected device was not returned.A post-operative x-ray picture was received.It could be confirmed that the plate has dislocated and that two screws have backed out.Despite confirming the event, it was not possible to determine the exact root cause.In addition to the return of the devices, more information such as patient medical records and patient post-operative behavior as well as more x-ray images, would have been required for a complete investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|