• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. NEXGEN FEMORAL IMPACTOR HEAD; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. NEXGEN FEMORAL IMPACTOR HEAD; INSTRUMENT, KNEE Back to Search Results
Catalog Number 00590103200
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during an initial total knee arthroplasty, the femoral impactor head fractured upon impaction of the implant.There was no patient impact and no adverse events have been reported as a result of the malfunction.All available information has been provided.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d4; g3; h2; h3; h4; h5; h6; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN FEMORAL IMPACTOR HEAD
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23189750
Report Number0001822565-2025-03617
Device Sequence Number9843453
Product Code LXH
UDI-Device Identifier00889024201668
UDI-Public(01)00889024201668(10)ZB6803160
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00590103200
Device Lot NumberZB6803160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2025
Supplement Date Received by Manufacturer01/05/2026
Initial Report FDA Received Date10/01/2025
Supplement Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-