| Catalog Number |
SRL-320 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side ¿rupture¿.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, b6, d6(b), h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: not observed through visual inspection.None of the other observations performed during the device analysis (creases and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿rupture¿.This record is for the right side.Device was explanted.Device will be returned.
|
| |
|
Search Alerts/Recalls
|