• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH RONDO; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH RONDO; EXTERNAL COMPONENT Back to Search Results
Model Number RONDO 3 AUDIO PROCESSOR [ME1550]
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event or Problem Description
The device was sent to service repair for repair.Upon preliminary investigation a broken and leaking battery was found.
 
Additional Manufacturer Narrative
Conclusion: during the repair process of a rondo 3 audio processor, a leaking battery was detected.Due to the destroyed device housing, it can be assumed that the damage to the rechargeable battery inside is caused by strong mechanical stress.There has been no report of patient injury from contact with leaking electrolyte.This is a final report.
 
Event or Problem Description
The device was sent to service repair for repair.Upon preliminary investigation a broken and leaking battery was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RONDO
Common Device Name
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key23199068
Report Number9710014-2025-00926
Device Sequence Number10061467
Product Code MCM
UDI-Device Identifier09008737369644
UDI-Public(01)09008737369644
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberRONDO 3 AUDIO PROCESSOR [ME1550]
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer09/29/2025
Initial Report FDA Received Date10/02/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-