|
|
| Catalog Number |
EPM8755 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).H3 photo investigation summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photos show a sales unit box with product codes epm8755, lot#: 103j0m, expiration date: 2029-08-31, packaging information without product inside.Based on the photo review, the event reported is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.H6 component code: g07002 - device evaluation pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a coronary artery bypass procedure on (b)(6) 2025 and suture was used.During the procedure, it was reported that the problem of pulling off suture needle happened when suturing blood vessels.Total 3 products occurred needle pull off issue.Changed another one to complete the surgery.There is no report on patient's injury.There will return 1 actual products and 4 representative samples for analysis.Enclosed please find defect photo.No additional information could be provided.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: d4 udi number.Additional h6 codes.Additional information was requested, and the following was obtained: received information as following from sales rep via phone today.After contacting the hospital, the hospital feedback that the surgery was coronary artery bypass grafting.There were 3 consecutive needle pull-off issues with the same lot of products during suturing during the surgery.Then the surgeon replaced a new suture (with specific product information unknown) to continue the suturing.The subsequent surgery went smoothly.Vascular damage (specific details unknown) and prolonged surgical time (specific extension time unknown) due to repeated puncture in this event.No subsequent adverse event report was received.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.One detached needle and one small needle- suture were received for analysis.Product code epm8755, this product code is double armed.Upon visual inspection of the detached needle, the swage and attachment area were noted to be as expected.The barrel hole was examined under magnification, and no suture remnant was noted.In the inspection of the needle-suture piece, the swage and attachment area were noted to be as expected, and no needle pull off was noted.The suture was examined, and the extreme was noted to be cut probably caused by a surgical instrument.The functional test was not possible because the suture was received with a short length.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached as to what caused the reported complaint.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.Four representative unopened samples were received for analysis.Product code epm8755.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.
|
| |
|
Search Alerts/Recalls
|
|
|