• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INVIVE?; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INVIVE?; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V173
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited an alert message indicating that battery replacement indicator had been reached on (b)(6) 2014.Technical services discussed that this was associated with unexpected hardware resets and the device data was consistent with a high impedance battery and recommended device replacement.This device was subsequently explanted and was not returned.Other than the surgical procedure, no additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited an alert message indicating that battery replacement indicator had been reached on 17jul2014.Technical services discussed that this was associated with unexpected hardware resets and the device data was consistent with a high impedance battery and recommended device replacement.This device was subsequently explanted and was not returned.Other than the surgical procedure, no additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device determined it had undergone resets and that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption (during telemetry or other higher power device operations) and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE?
Common Device Name
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23200736
Report Number2124215-2025-69754
Device Sequence Number13282137
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public(01)00802526536632(17)160508
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/08/2016
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number106344
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2025
Supplement Date Received by Manufacturer04/03/2026
Initial Report FDA Received Date10/02/2025
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2016-2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age87 YR
Patient SexMale
-
-