| Model Number |
M0066703081 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Peeled/Delaminated (1454); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Approximated based on the date the manufacturer became aware of the event, as no event date was reported.Imdrf device code a24 captures the reportable investigation results of sleeve detachment of device or device component.The sensor wire was returned for analysis.The visual and microscopic analysis was performed, and it was observed that the sleeve was detached from the wire and the polytetrafluoroethylene (ptfe) was in good condition.With all the available information, boston scientific concludes that the reported event of ptfe peeled was not confirmed.Based on the analysis of returned device, it is possible to conclude that operational factors such as some manipulation during the preparation of the device could have caused the sleeve detached from the wire.Based on the information available and investigation results, an investigation conclusion code of adverse event related to procedure was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, the distal tip peeled off.The procedure was completed with another of the same device without patient complications.Analysis of the returned guidewire identified the sleeve was detached from the guidewire.
|
| |
|