| Model Number |
RONDO 3 AUDIO PROCESSOR [ME1550] |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device was sent to service repair for repair.Upon preliminary investigation a broken and leaking battery was found.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: during the repair process of a rondo 3 audio processor, a leaking battery was detected.Due to the destroyed device housing, it can be assumed that the damage to the rechargeable battery inside is caused by strong mechanical stress.There has been no report of patient injury from contact with leaking electrolyte.This is a final report.
|
| |
|
Event or Problem Description
|
|
The device was sent to service repair for repair.Upon preliminary investigation a broken and leaking battery was found.
|
| |
|
Search Alerts/Recalls
|