Model Number ABSTACK30X
Medical Device Problem Codes Positioning Failure (1158); Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, intra-operatively, all four tacks fired did not adhere to the tissue, then the fifth tack fired jammed.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 correction: d4 (unique identifier (udi) #) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted noted one instrument returned with the blister pack.Tacks were observed within the tube and were not protruding.During functional testing, the handle was actuated, and the instrument cycled properly.The remaining twenty two tacks deployed and seated properly in the test media.No functional abnormalities were observed.It was reported that tack did not hold and the tacker did not fire.The reported issues were not confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: bent shaft.Visual inspection noted that the tube assembly was observed to be bent and kinked.The product analysis noted evidence that the device was not used as intended.This issue can occur if the instrument is applied with excessive force or side load, causing excess torque on the rotating helices and tube shaft which can cause poor tack penetration.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: appropriate force should be applied to the handle of the device when placing tacks.Excessive force may result in damage to the tissue, device and/or the material being fixated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORBATACK
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23208183
Report Number1219930-2025-04419
Device Sequence Number9139453
Product Code GDW
UDI-Device Identifier10884521184213
UDI-Public(01)10884521184213(17)280430(10)N5E1718Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123109
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberABSTACK30X
Device Catalogue NumberABSTACK30X
Device Lot NumberN5E1718Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2025
Supplement Date Received by Manufacturer11/21/2025
Initial Report FDA Received Date10/03/2025
Supplement Report FDA Received Date12/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-