| Model Number |
ABSTACK30X |
| Medical Device Problem Codes |
Positioning Failure (1158); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, all four tacks fired did not adhere to the tissue, then the fifth tack fired jammed.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 correction: d4 (unique identifier (udi) #) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted noted one instrument returned with the blister pack.Tacks were observed within the tube and were not protruding.During functional testing, the handle was actuated, and the instrument cycled properly.The remaining twenty two tacks deployed and seated properly in the test media.No functional abnormalities were observed.It was reported that tack did not hold and the tacker did not fire.The reported issues were not confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: bent shaft.Visual inspection noted that the tube assembly was observed to be bent and kinked.The product analysis noted evidence that the device was not used as intended.This issue can occur if the instrument is applied with excessive force or side load, causing excess torque on the rotating helices and tube shaft which can cause poor tack penetration.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: appropriate force should be applied to the handle of the device when placing tacks.Excessive force may result in damage to the tissue, device and/or the material being fixated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|