| Catalog Number |
150NSS35 |
| Medical Device Problem Codes |
Material Protrusion/Extrusion (2979); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Exposure to Body Fluids (1745)
|
| Date of Event |
09/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that, after the guidewire was used to guide the ureteroscope into the ureter, the guidewire was withdrawn.The surface of the guidewire became wrinkled, rendering it unable to guide stent placement.Patient has been exposed to harmful substances, blood or body fluids.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: f, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that, after the guidewire was used to guide the ureteroscope into the ureter, the guidewire was withdrawn.The surface of the guidewire became wrinkled, rendering it unable to guide stent placement.Patient has been exposed to harmful substances, blood or body fluids.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.Visual evaluation of the returned one-photo sample noted one opened (with original packaging), nitinol guidewire.Visual inspection of the single photo sample indicated that the nitinol guidewire was present inside the packaging, with the tip protruding through the packaging with missing coating.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that, after the guidewire was used to guide the ureteroscope into the ureter, the guidewire was withdrawn.The surface of the guidewire became wrinkled, rendering it unable to guide stent placement.Patient has been exposed to harmful substances, blood or body fluids.No medical intervention was reported.
|
| |
|