MAUDE Adverse Event Report: AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 24FR DILATOR; DH EF PERC PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 24FR DILATOR; DH EF PERC PLACEMENT PRODUCTS Back to Search Results
Model Number 98434
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record for the reported lot number, 30355150, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The actual complaint product was not returned for evaluation.All information reasonably known as of (b)(6) 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported that the sutures broke 48-hours after insertion."no incidents during insertion, simple procedure.Experienced professional, familiar equipment.On (b)(6) 2025 a 3-stage computed tomography (ct) with intravenous (iv) was performed and it noted: -at the abdominopelvic level: gastrostomy tube in intragastric position.-tubes removed with distal metal part in intragastric position on a dependent slope.-no abscess or bleeding lesion identified today.-no focal parietal thickening of the digestive loops within the limits of the examination.-non-specific pneumoperitoneum following the insertion of a gastrostomy tube.A small amount.Actions taken in the healthcare facility for the patient's care included "patient monitoring strengthened.".
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30355150, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.Root cause could not be determined.All information reasonably known as of 24-oct-2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 24FR DILATOR
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key23211193
Report Number9611594-2025-00228
Device Sequence Number18386137
Product Code KGC
UDI-Device Identifier00350770984346
UDI-Public(01)00350770984346(240)109843404(11)250122(17)261127(10)30355150
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number98434
Device Catalogue Number98434
Device Lot Number30355150
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2025
Supplement Date Received by Manufacturer09/24/2025
Initial Report FDA Received Date10/03/2025
Supplement Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
-
-