| Model Number |
1013731 |
| Medical Device Problem Codes |
Break (1069); Peeled/Delaminated (1454); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the distal superficial femoral artery, popliteal, and posterior tibial artery.Following pre-dilation of the vessel with an unknown balloon, three 10cm ht command 18 guidewires (gw) were attempted to be advanced and were noted to have difficulty when being inserted into the 5 french micro sheath, and the coating became bunched at the tip.Therefore, a new command wire was used and advanced through the same sheath successfully and the procedure was completed.There was no adverse patient effect and no clinically significant delay in the procedure.One device was received and return analysis revealed that the polymer coating was separated, torn and folded over itself.Additional information was received that the reported coating becoming bunched at the tip meant that the polymer coating was separated and folded over itself with tear being noted.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional devices referenced in b5 are filed under separate medwatch numbers.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection and dimensional analysis were performed on the returned device.The reported polymer damages were confirmed.The reported difficulty to advance was not confirmed as it cannot be replicated in a testing environment and due device condition.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided and the observations from the returned analysis, a definitive cause for the reported issue could not be determined.Factors that may contribute to difficulty advancing the guide wire into a sheath include, but are not limited to, inner diameter of the sheath, device support, buildup of procedural contaminants, challenging anatomy, user technique, and/or damage to the sheath or guide wire which likely led to the reported damages to the polymer coating.Based on the results of the complaint investigation, there is no indication a product quality issue with respect to manufacture, design or labeling of the device.H6 correction: medical device problem code 1454 removed.
|
| |
|